Publication: HPLC and LC-MS/MS methods for determination of sodium benzoate and potassium sorbate in food and beverages: Performances of local accredited laboratories via proficiency tests in Turkey
Program
Authors
Authors
Goren, AHMET CEYHAN
BILSEL, Gokhan
SIMSEK, Adnan
BILSEL, Mine
AKCADAG, Fatma
TOPAL, Keyser
OZGEN, Hasan
Advisor
Language
Type
Publisher
Journal Title
Journal ISSN
Volume Title
Abstract
High Performance Liquid Chromatography LC-UV and LC-MS/MS methods were developed and validated for quantitative analyses of sodium benzoate and potassium sorbate in foods and beverages. HPLC-UV and LC-MS/MS methods were compared for quantitative analyses of sodium benzoate and potassium sorbate in a representative ketchup sample. Optimisation of the methods enabled the chromatographic separation of the analytes in less than 4 min. A correlation coefficient of 0.999 was achieved over the measured calibration range for both compounds and methods (HPLC and LC-MS/MS). The uncertainty values of sodium benzoate and potassium sorbate were found as 0.199 and 0.150 mg/L by HPLC and 0.072 and 0.044 mg/L by LC-MS/MS, respectively.
Description
Source:
Keywords:
Keywords
Citation
Goren A. C. , BILSEL G., SIMSEK A., BILSEL M., AKCADAG F., TOPAL K., OZGEN H., -HPLC and LC-MS/MS methods for determination of sodium benzoate and potassium sorbate in food and beverages: Performances of local accredited laboratories via proficiency tests in Turkey-, FOOD CHEMISTRY, cilt.175, ss.273-279, 2015