Rapid and reliable 25-OH vitamin D2 and 25-OH vitamin D3 measurements by multitasker LC-MS/MS

No Thumbnail Available
Date
2018-01-01T00:00:00Z
Authors
Goren, AHMET CEYHAN
SAHIN, Gozde
GUMILCINELI, Ismail
BINICI, Burcu
Journal Title
Journal ISSN
Volume Title
Publisher
Research Projects
Organizational Units
Journal Issue

Metrics

Search on Google Scholar

Abstract
The 25-OH Vitamin D2 and 25-OH Vitamin D3 measurement method was validated using a Zivak Multitasker Fully Automated Sample Preparation and Injection System coupled to a MS detector. The current method incorporates all sample preparations, SPE clean up procedures, LC and MS facilities into an all in one system. This system can report patient data within 6 minutes including the SPE and Sample preparation and mass spectrometry analysis time. The correlation coefficient of the developed method was greater than 0.99 for measurands, 25-OH-vitamin Ds, in the calibration range. The measurement uncertainty of the measurements were evaluated and found as 7.7 % and 7.6 % for 25-OH vitamin D2 and 25-OH-vitamin D3, respectively.
Description
Keywords
Citation
Goren A. C. , SAHIN G., GUMILCINELI I., BINICI B., -Rapid and reliable 25-OH vitamin D2 and 25-OH vitamin D3 measurements by multitasker LC-MS/MS-, JOURNAL OF CHEMICAL METROLOGY, cilt.12, ss.17-25, 2018