Publication:
Effects of High-Energy Curing Lights on Time-Dependent Temperature Changes of Pulp Space During Orthodontic Bonding

dc.contributor.authorGÜL AMUK, NİSA
dc.contributor.authorKURT, GÖKMEN
dc.contributor.authorER, ÖZGÜR
dc.contributor.authorÇAKMAK, GÜLŞEN
dc.contributor.authorASLANTAŞ, VEYSEL
dc.contributor.institutionauthorKURT, GÖKMEN
dc.date.accessioned2020-01-15T21:39:03Z
dc.date.available2020-01-15T21:39:03Z
dc.date.issued2019-03-01T00:00:00Z
dc.description.abstractObjective: The purpose of this study was to assess the temperature changes and cooling times during orthodontic bonding by a light-emitting diode (LED) and plasma arc lights (PAC) in different time and power modes with thermal imaging. Methods: A total of 100 human permanent upper first premolar teeth were included in the study. Five groups were conducted, 20 teeth each, and different energy outputs of curing lights were used for adhesive polymerization with different exposure times. The temperature changes in the pulp space and cooldown times were measured by a thermal imaging system. A paired t-test, analysis of variance (ANOVA), and Student-Newman-Keuls multiple comparison tests were used for data analysis. Results: A statistically significant temperature rise was detected with all curing lights (p<0.05). The greatest temperature changes were observed in the LED standard mode with 10 seconds of exposure time (6.66±1.98°C) and LED extra power mode with 6 seconds of exposure time (6.50±1.64°C) among groups, while using PAC for 3 seconds created the smallest temperature increase (1.81±0.99°C). An application of the LED extra power mode for 6 seconds exhibited the longest cooldown time (205.91±47.48 seconds), and the shortest cooldown time was detected as 71.30±43.15 seconds with the PAC 3-second application. Conclusions: LED lights with an increased exposure time induced significant temperature rises, while no PAC light group exceeded the critical threshold value. The exposure time is more important than the energy output level of the light-curing system on temperature increments of the pulp chamber.en
dc.identifier.citationGÜL AMUK N., KURT G., ER Ö., ÇAKMAK G., ASLANTAŞ V., -Effects of High-Energy Curing Lights on Time-Dependent Temperature Changes of Pulp Space During Orthodontic Bonding-, Turkish Journal of Orthodontics, cilt.32, ss.22-27, 2019
dc.identifier.doi10.5152/turkjorthod.2019.19015
dc.identifier.pubmed30944896
dc.identifier.scopus85069766770
dc.identifier.urihttps://openaccess.bezmialem.edu.tr/handle/20.500.12645/13268
dc.identifier.wosWOS:000462161300005
dc.identifier.yoksis5259944
dc.language.isoen
dc.rightsinfo:eu-repo/semantics/openAccessen
dc.titleEffects of High-Energy Curing Lights on Time-Dependent Temperature Changes of Pulp Space During Orthodontic Bonding
dc.typeArticle
dspace.entity.typePublication
local.avesis.id7e2baf14-4d42-414b-9d1b-5b89a46647eb
local.publication.isinternational0
relation.isAuthorOfPublication3e64bd84-5cc2-4d20-a9d6-bbda78ee6d9f
relation.isAuthorOfPublication.latestForDiscovery3e64bd84-5cc2-4d20-a9d6-bbda78ee6d9f
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
107-10.5152-TurkJOrthod.2019.19015.pdf
Size:
839.98 KB
Format:
Adobe Portable Document Format
Description: