In vitro comparison of working length determination using three different electronic apex locators.

dc.contributor.authorKuştarci, Alper
dc.contributor.authorArslan, Dilara
dc.contributor.authorAltunbaş, Demet
dc.date.accessioned2023-05-16T16:38:29Z
dc.date.available2023-05-16T16:38:29Z
dc.date.issued2014-08-31T21:00:00Z
dc.description.abstractThe aim of this study was to compare the accuracy of the apex-locating functions of DentaPort ZX, Raypex 5 and Endo Master electronic apex locators (EALs) in vitro.
dc.description.abstractThirty extracted human single-rooted teeth with mature apices were used for the study. The real working length (RWL) was established by subtracting 0.5 mm from the actual root canal length. All teeth were mounted in an alginate model that was especially developed to test the EALs and the teeth were then measured with each EAL. The results were compared with the corresponding RWL, which was subtracted from the electronically determined distance. Data were analyzed using a paired-samples t-test, a Chi-square test and a repeated measure analysis of variance evaluation at the 0.05 level of significance.
dc.description.abstractStatistical analysis showed that no significant difference was found among all EALs (P > 0.05).
dc.description.abstractThe accuracy of the EALs was evaluated and all of the devices showed an acceptable determination of electronic working length between the ranges of ±0.5 mm.
dc.identifier.pubmed25426148
dc.identifier.urihttps://hdl.handle.net/20.500.12645/38189
dc.language.isoen
dc.subjectDentaPort ZX
dc.subjectEndo Master
dc.subjectRaypex 5
dc.subjectelectronic apex locator
dc.subjectworking length
dc.titleIn vitro comparison of working length determination using three different electronic apex locators.